Ultrasonic milkanalyser
Operation manual
86/92
5. Co Meter En/Dis.
Enables or disables the conductivity measurement system. The following
message appears on the display:
Cond Measuring
Yes
No OK
Yes
6.Corrections in conductivity measurement
It is done by starting the menu
Corrections -> Cond measure.
You have the
possibility to increase/decrease the measured conductivity value from
– 1.00
till +1.00, with step 0.01. After starting this function the display shows the
following:
Con Meter
-1.0<=Corr>=1.0
Co Corr=+0.00
Edit
– Up/Down
The cursor is positioned below the +. By using buttons
Up/Down
, the
operator has the possibility to change the value (number). By pressing the
button
Enter
, the operator confirms the chosen value and moves to the next
position for editing it. After the last position is edited, if the correction value is
within allowed boundaries, the following is displayed:
Co Corr Saved
, which
means, that the correction is entered and saved. On the contrary
– it returns
at the beginning and expects valid correction.
7. Conductivity calibration buffer preparation
In order a standard buffer for conductivity measuring to be prepared follow
the instruction below:
1. Take the packet with the powder buffer.
2. Carefully shake the packet in order to gather the powder at the bottom.
3. Cut one end of the packet.
Содержание MIA-LA
Страница 1: ...Operation manual 1 92 MIA LA MILK ANALYZER LCD display 4 lines x 16 characters Operation manual...
Страница 12: ...Ultrasonic milkanalyser Operation manual 12 92 Fig 3 Principle working scheme...
Страница 13: ...Ultrasonic milkanalyser Operation manual 13 92 Fig 4 Connecting peripheral devices...