Ultrasonic milkanalyser
Operation manual
24/92
Cleaning
Please wait
If there is a glass with sample it is completely emptied and then the analyser
automatically fulfills procedure for sucking the cleaning solution and c rinsing
the analyser's inner system. In order the input metal pipe to be cleaned out
for a short time a cleaning solution is pumped out for a short time in the
already empty glass or in case there is no empty glass there
– in the funnel
for the waste liquids. It is filled to the middle and then is sucked back and
poured in the reservoir with the used liquids. In case that you are sure that
you’ll not use the analyser for a longer period, but it will be connected to the
power supply and switched on, which means that the process of automatic
clean will be active, do not leave sample holder in the recess of the analyser,
in order not to be overfilled during cleaning from the output pipe. Then the
cleaning solution will be directly poured in the funnel for waste liquids.
After the cleaning is finished the displays shows the following:
End of cleaning
After 2 seconds the display shows:
Analyser ready
Then the analyser is ready for normal measurement.
4.1.1.2. Manual start
The current cleaning may be completed by manual start of the menu
Cleaning
. It is used before starting the menu Final Clean. It serves for
cleaning the fats from the measuring tract with alkaline cleaning solution with
which the canister is filled in. It is started using the menu for choosing the
working mode of the analyser. After the measurement is completes, by
continuous pressing the button
Enter
, the possible analyser operation modes
are shown on the display.