Ultrasonic milkanalyser
Operation manual
25/92
Cow
Sheep
UHT
Cleaning
Final Clean
By choosing
Cleaning
the current cleaning is started. The display shows the
following:
Auto clean
started!
Put empty glass
In this way the current cleaning is started. The procedure and operator’s
actions are described above.
Do not switch off the device at the end of the working day before the
automatic cleaning procedure is completed. If it is not automatically cleaned
and there is not a possibility to wait starting the automatic cleaning, then start
manually the cleaning procedure with alkaline cleaning solution and if it is
necessary with acidic cleaning solution as it is described below.
4.1.2. End (final) cleaning.
4.1.2.1. Final cleaning.
It is done with 3% acidic cleaning detergent (for working solution preparation
see 4.2.2.). It is done daily. Serves for cleaning the protein deposits from the
measuring system of the analyser, which were not removed during work with
the alkaline cleaning solution and removing milk stone layers.
Do not use chemicals, which are not intended for milking equipment and
vessels in dairy industry. Pay particular attention to the acidic cleaning
solution concentration. The higher concentration may damage the measuring
sensor.