Ultrasonic milkanalyser
Operation manual
49/92
There are the following possibilities:
With button
Exit
– to leave the menu.
With button
Yes
– to confirm the chosen from the list calibration name.Now the
program returns to the beginning of the menu for setting calibration names.
With button
Next
– to show the next calibration name from the list.
If it is decided a new calibration name to be edited, the display shows:
Cal1: Sheep
User Edited
Name:
Prev Set Next
There are the following possibilities:
With button
Prev
– to display the previous ASCII symbol.
With button
Set
– to confirm the ASCII symbol, shown on the display and
passes to editing the next symbol from the calibration name.
With button
Next
– to show the next ASCII symbol.
After editing the last (eighth) name symbol, the display shows:
Cal1: Sheep
User Edited
Name:MilkShp
Exit Save
There are the following possibilities:
With button
Exit
– to leave the menu.
With button
Save
– to confirm already edited calibration name and to save it in
the device. The program returns to the beginning of the menu for setting
calibration names.
7.2.3.5. Set Base FrPnt
Through this menu you have the possibility to enter the basic freezing point
separately for each calibration. For more information, see
Appendix Freezing
point.
After choosing the menu the following is displayed: