Current calibration content is not changed, the analyser continues using it,
but there is a reserve copy in an internal buffer.
The current calibration is replaced with the calibration from the internal buffer
and the analyser starts working with it. The content of the internal buffer is not
changed.
If after recalibration “Save calibration” is pressed the new calibration
settings will be saved over the factory settings. After that is impossible
to restore the factory settings of the calibration. Save the newly made
calibration only if you are sure about its correctness.
7.2.3.4. Settings Page 2.
After this menu is started the display shows the following:
Settings Page2
Set Calibr Name
Select High Fat
HFSpeed for Cal
Result Precision
PCB Main Identi
Larg Res En/Dis
Exit
Now there is a possibility one of the following options to be set:
7.2.3.4.1.Larg Res En/Dis.
The format of the measurement data send towards the computer is set.
Current working
calibration (1, 2 or 3)
Save
Internal storing
buffer
Current working
calibration (1, 2 or 3)
Restore
Internal storing
buffer