Setup Messages
Chapter Five: Remote RS-232 Operation
82
To check the sensor range of your unit, issue the request:
R 55
for the Low range sensor
R 33
for the High range sensor
The controller responds with the message [
E
x
value
], where x corresponds to either ‘H’ or
‘L’ for High or Low sensor range and
value
corresponds to a valid sensor range shown in Table
14, page 81.
An example response, if the Low range sensor has a range of 100 Torr, is:
EL
08
To change the Low range to 5 Torr, enter:
EL
05
To change the High range to 1000 Torr, enter:
EH
10