Chapter Five: Remote RS-232 Operation
Setup Messages
81
How To Set the Sensor Range
Note
Be sure that the sensor is connected to the 651 controller before changing
the sensor range.
The command [
E
x
value
] identifies the range of the sensor, in units, where
x
corresponds to
either ‘H’ or ‘L’ for the High or Low pressure sensor, and
value
corresponds to a valid sensor
range, as listed in Table 14.
The 651 controller is initially configured to work with a 10 and a 1000 Torr pressure sensor. If
your sensors cover different pressure ranges, use this command to identify the range of your
sensor.
Sensor Range Values
Value
Sensor Range
Value
Sensor Range
00
0.1 Torr
10
1000 Torr
01
0.2
11
5000
02
0.5
12
10000
03
1
13
1.33 mBar
04
2
14
2.66
05
5
15
13.33
06
10
16
133.3
07
50
17
1333
08
100 (initial)
18
6666
09
500
19
13332
21
20
22
200
Table 14: Sensor Range Values
Note
1. Pressure readings are reported as a percentage (%) of full scale (FS),
where full scale is the sensor range shown in Table 14.
2. For example, if the actual pressure is 10 Torr for a 10 Torr FS unit,
the 651 controller reports a pressure value of 100 (for 100%). If the
pressure is 10 Torr for a 100 Torr FS unit, the 651 controller reports
a pressure value of 10 (for 10%).
Содержание 651D Series
Страница 1: ...118965 P1 Preliminary 5 02 MKS Type 651D Pressure Controller...
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 20: ...Customer Support Chapter One General Information 8 This page intentionally left blank...
Страница 50: ...Labels Chapter Three Overview 38 This page intentionally left blank...
Страница 78: ......
Страница 107: ...Chapter Five Remote RS 232 Operation Setup Messages 95 To change the softstart rate for set point A to 50 enter I 1 50...
Страница 130: ...How To Activate the Softstart Rate Chapter Six Remote Digital Logic Control 118 This page intentionally left blank...
Страница 142: ...Troubleshooting Chapter Seven Maintenance and Troubleshooting 130 This page intentionally left blank...
Страница 149: ...Appendix C Initial Settings 137 This page intentionally left blank...
Страница 150: ......
Страница 158: ...Appendix E RS 232 Message Summary 146 This page intentionally left blank...