Setup Messages
Chapter Five: Remote RS-232 Operation
80
How To Set the Sensor Signal Input Range
Note
Be sure that the sensor is connected to the 651 controller before changing
the sensor signal input range.
The command [
G
value
] sets the full scale voltage range for the pressure sensor input,
where:
value
:
0
= 1 Volt
1
= 5 Volts
2
= 10 Volts (initial)
To query the current range of the sensor signal input, issue the request:
R 35
The controller responds with the message [
G
value
], where:
value
: 0
= 1 Volt
1
= 5 Volts
2
= 10 Volts
An example response, if the sensor input has a full scale range of 10 Volts, is:
G
2
To change the sensor signal input range to 5 Volts, enter:
G
1
Содержание 651D Series
Страница 1: ...118965 P1 Preliminary 5 02 MKS Type 651D Pressure Controller...
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 20: ...Customer Support Chapter One General Information 8 This page intentionally left blank...
Страница 50: ...Labels Chapter Three Overview 38 This page intentionally left blank...
Страница 78: ......
Страница 107: ...Chapter Five Remote RS 232 Operation Setup Messages 95 To change the softstart rate for set point A to 50 enter I 1 50...
Страница 130: ...How To Activate the Softstart Rate Chapter Six Remote Digital Logic Control 118 This page intentionally left blank...
Страница 142: ...Troubleshooting Chapter Seven Maintenance and Troubleshooting 130 This page intentionally left blank...
Страница 149: ...Appendix C Initial Settings 137 This page intentionally left blank...
Страница 150: ......
Страница 158: ...Appendix E RS 232 Message Summary 146 This page intentionally left blank...