Operation
Chapter Four: Local Operation
64
4. Press the [L
EARN
] and [Z
ERO
] keys simultaneously for about three seconds.
The 651 controller
learns
the input voltage that corresponds to the analog set point zero
value, then the screen returns to the default
Pressure and Position
display screen.
How To Calibrate the Full Scale of the Analog Set Point
The analog set point can be configured for 5 V (initial) or 10 V full scale input (refer to step 9).
You can operate the 651 controller, however, at a different full scale input, for example, 4.5 V,
by recalibrating the full scale of the analog set point. Although the actual voltage input to the
unit must be measured, you can report the voltage input applied as a percentage of the analog set
point full scale, using the analog set point display.
For example, if the analog set point is configured for 5 V full scale input, and the actual input
applied is 5 V, the analog set point display shows a value of 100%. However, if the actual input
applied is 2.5 V, the analog set point display indicates that input by showing a value of 50%.
To recalibrate the full scale of the analog set point:
1. Be sure that the default
Pressure and Position
display appears on the screen.
2. Apply the full scale input voltage (for example, 4.5 V) to pins 33 and 34 on the Interface
connector. (Pin 34 should be connected to Pin 35 to properly ground the analog input.)
Refer to Table 8, for the Interface connector pinout.
Note
Pin 34 should be connected to Pin 35 to properly ground the analog
input.
3. Press the [
] and [S
TOP
] keys simultaneously for about three seconds.
The screen displays:
CAUTION
CALIBRATION MODE
4. Press the [S
ETPOINT
D] key.
The screen displays:
ANLG SP:
95.00 %
FULL SPAN: 20015
Содержание 651D Series
Страница 1: ...118965 P1 Preliminary 5 02 MKS Type 651D Pressure Controller...
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 20: ...Customer Support Chapter One General Information 8 This page intentionally left blank...
Страница 50: ...Labels Chapter Three Overview 38 This page intentionally left blank...
Страница 78: ......
Страница 107: ...Chapter Five Remote RS 232 Operation Setup Messages 95 To change the softstart rate for set point A to 50 enter I 1 50...
Страница 130: ...How To Activate the Softstart Rate Chapter Six Remote Digital Logic Control 118 This page intentionally left blank...
Страница 142: ...Troubleshooting Chapter Seven Maintenance and Troubleshooting 130 This page intentionally left blank...
Страница 149: ...Appendix C Initial Settings 137 This page intentionally left blank...
Страница 150: ......
Страница 158: ...Appendix E RS 232 Message Summary 146 This page intentionally left blank...