Setup Messages
Chapter Five: Remote RS-232 Operation
90
How To Set the Gain Values
The command [
M
x
value
] sets the gain values for the internal set points, where:
x
:
1
= Set point A gain
2
= Set point B gain
3
= Set point C gain
4
= Set point D gain
5
= Set point E gain
value
:
0 to 100%
To set the gain for set point A to 50%, enter:
M
1 50
To report the gain value for any set point, issue the request:
R
xx
where
xx
:
46 = Set point A gain
47 = Set point B gain
48 = Set point C gain
49 = Set point D gain
50 = Set point E gain
To report the gain value for set point A, enter:
R 46
The controller responds with the message [
M
x value
], where:
x
:
1
= Set point A gain
2
= Set point B gain
3
= Set point C gain
4
= Set point D gain
5
= Set point E gain
value:
0 to 100%
An example response, if the gain for set point A is set to 45, is:
M
1 45
Содержание 651D Series
Страница 1: ...118965 P1 Preliminary 5 02 MKS Type 651D Pressure Controller...
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 20: ...Customer Support Chapter One General Information 8 This page intentionally left blank...
Страница 50: ...Labels Chapter Three Overview 38 This page intentionally left blank...
Страница 78: ......
Страница 107: ...Chapter Five Remote RS 232 Operation Setup Messages 95 To change the softstart rate for set point A to 50 enter I 1 50...
Страница 130: ...How To Activate the Softstart Rate Chapter Six Remote Digital Logic Control 118 This page intentionally left blank...
Страница 142: ...Troubleshooting Chapter Seven Maintenance and Troubleshooting 130 This page intentionally left blank...
Страница 149: ...Appendix C Initial Settings 137 This page intentionally left blank...
Страница 150: ......
Страница 158: ...Appendix E RS 232 Message Summary 146 This page intentionally left blank...