Chapter Six: Remote Digital Logic Control
How To Activate the Softstart Rate
117
How To Activate the Softstart Rate
When using the front panel controls or RS-232 to Open the valve, Close the valve, or use a
setpoint, the softstart rate is automatically activated. Having a softstart rate setting of 100% is the
same as not applying any softstart rate.
The following section applies to the use of TTL commands to Open the valve, Close the valve,
or use a setpoint.
The softstart rate for the active set point is controlled by the state of digital logic input pin 7 on
the Interface connector. This line must be held low to activate the softstart rate.
To activate softstart control of the active set point:
1. Hold the softstart line low (pin 7 on the Interface connector).
Refer to Table 8, for the Interface connector pinout. Hold the TTL low signal for a
minimum of 50 milliseconds. If the line is
not
held low, the valve will move at 100%
full speed.
Содержание 651D Series
Страница 1: ...118965 P1 Preliminary 5 02 MKS Type 651D Pressure Controller...
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 20: ...Customer Support Chapter One General Information 8 This page intentionally left blank...
Страница 50: ...Labels Chapter Three Overview 38 This page intentionally left blank...
Страница 78: ......
Страница 107: ...Chapter Five Remote RS 232 Operation Setup Messages 95 To change the softstart rate for set point A to 50 enter I 1 50...
Страница 130: ...How To Activate the Softstart Rate Chapter Six Remote Digital Logic Control 118 This page intentionally left blank...
Страница 142: ...Troubleshooting Chapter Seven Maintenance and Troubleshooting 130 This page intentionally left blank...
Страница 149: ...Appendix C Initial Settings 137 This page intentionally left blank...
Страница 150: ......
Страница 158: ...Appendix E RS 232 Message Summary 146 This page intentionally left blank...