56
8-5. Key lock function
Lock the key function limit a key operation of front panel for avoids wrong operation of keys.
8-5-1. Key lock 1
The function of F-08 sets whether the key operation of [ON/OFF], [SET], [ACCUM./ ] and [PRESET
TARE/ ] is [valid] or [invalid].
Setting range: 0000-1111, 0 : Invalid, 1 : Valid
10
0
digit : ON/OFF
10
1
digit : SET
10
2
digit : ACCUM./
10
3
digit : PRESET TARE/
The operation of [SET] key becomes valid only once by pressing [SET] key and [NET/GROSS] key at
the same time when the operation of [SET] key is locked.
8-5-2. Key lock 2
The function of F-09 sets whether the key operation of [TARE/
], [NET/GROSS/
], [ZERO] or [F/
] is [valid] or [invalid].
Setting range: 0000-1111, 0: Invalid, 1: Valid
10
0
digit: TARE/
10
1
digit: NET/GROSS/
10
2
digit: ZERO
10
3
digit: F/
8-6. Stability detection
When the change width of the load display is within the data width set by function F-10, and when
the condition continues for the time set by function F-11 or more, the stability detection is the
function to judge the measuring value being stable.
8-6-1. Data width of stability detection
The data width of stability detection can be set by function F-10.
Setting range : 000-999, Unit: 0.1D, 0
:
OFF
Default is set to [020].
The data width is obtained in the display conversion by the below formula.
[Data width of stability detection] = [Setting value of F-10] x 0.5 x [The scale interval]
When the setting of function F-10 is [5], and when the scale interval is [D = 2], it will become as follows;
[Data width stability detection] = 5 x 0.1 x 2 = 1D
8-6-2. Time width of stability detection
The time width of stability detection can be set by function F-11.
Setting range : 00-99, Unit : 0.1 s, 0
:
OFF
Default is set to [10].
When two range or three range switches have been selected by the automatic range switches, the data
width of the stability detection execute the display conversion of the data width by the value of the first
range scale interval (scale interval in the calibration).
Содержание CSD-903-EX
Страница 1: ...DIGITAL INDICATOR CSD 903 EX Instruction Manual EN294 1492 H ...
Страница 2: ...IV ...
Страница 91: ...75 ...
Страница 93: ...77 ...
Страница 95: ...79 ...
Страница 97: ...81 ...
Страница 129: ...113 14 4 Connecting method 14 4 1 One to one RS 485 RS 485 S G A B address 00 S G RS 485 host CSD 903 ...
Страница 199: ...183 Inform us about the contents of failure and situation at site in details ...
Страница 204: ...188 Inform us about the contents of failure and situation at site in details ...
Страница 206: ...190 Inform us about the contents of failure and situation at site in details ...
Страница 209: ...193 Inform us about the contents of failure and situation at site in details ...
Страница 224: ... ...
Страница 225: ......