195
RS-485 check
No.
Correct connecting?
Yes
Connect wires surely with an external equipment
according to the 16-3-11.
No.
The address number is
correctly set.
Yes
Switch RS-422/485 correctly.
No.
The communication
protocol is correctly set.
Yes
Set the flow control to “NONE”
from “YES”.
No.
Command/data is set properly.
Yes
Set the address number correctly according
to the 16-3-6.
No.
The host specifies the flow
control.
Yes
Set according to the specifications of host.
No.
The change of RS-422/485
is correctly set.
Yes
Use the correct command/data by referring
to the 16-3-12.
Inform Minebea about the contents of
failure and situation in details.
Yes
No.
There is a possibility not to be able to respond
because the host's response is slow. Enlarge
the response delay time of the instrument.
No.
Set the interval from the command receipt of host to
send is 1 ms or more.
The interval from
the command receipt of host to
send is 1 ms or more.
Is it appropriate
setting for the response delay
time to the response of host?
Yes
Inform us about the contents of
failure and situation at site in details
Содержание CSD-903-EX
Страница 1: ...DIGITAL INDICATOR CSD 903 EX Instruction Manual EN294 1492 H ...
Страница 2: ...IV ...
Страница 91: ...75 ...
Страница 93: ...77 ...
Страница 95: ...79 ...
Страница 97: ...81 ...
Страница 129: ...113 14 4 Connecting method 14 4 1 One to one RS 485 RS 485 S G A B address 00 S G RS 485 host CSD 903 ...
Страница 199: ...183 Inform us about the contents of failure and situation at site in details ...
Страница 204: ...188 Inform us about the contents of failure and situation at site in details ...
Страница 206: ...190 Inform us about the contents of failure and situation at site in details ...
Страница 209: ...193 Inform us about the contents of failure and situation at site in details ...
Страница 224: ... ...
Страница 225: ......