140
16-2. BCD output
16-2-1. Specifications of BCD output (P/No. : CSD903-P15)
Specifications Input/Output
Contents
BCD data
Output
8 digit parallel output.
POL. (Polarity)
Output
ON by minus polarity output, and OFF by plus polarity output.
P.C. (Print
command)
Output
Turning on during the fixed time after the conversion of BCD
output.
ERROR
Output
ON when the error occurs.
OVER
Output
ON synchronizing with the OL display (Over load).
Decimal point 10
1
,
Decimal point 10
2
,
Decimal point 10
3
Decimal point 10
4
Output
ON synchronizing with the decimal point display setting.
Stable
Output
ON when stability is detected.
Gross value
Output
ON when the gross data is output.
HOLD
Input
HOLD of BCD output.
BCD-ENABLE
Input
The BCD output is compulsorily put into the high impedance
condition.
16-2-2. Operation mode of BCD output
The operation mode of BCD output is selected by the setting of function F-30.
The operation mode of BCD output can be selected from [Stream mode], [Synchronizing with print],
[Synchronizing with accumulation] and [Synchronizing with finish].
Default is set to [Stream mode].
16-2-3. Target of BCD output
The target of BCD output is selected by the setting of function F-31.
Function No.
Set value
Contents
F-31
0 Display
interlock
1 Gross
value
2 Net
value
3 Tare
value
4
Accumulated total value
5 Accumulated
count
Default is set to [Display interlock].
16-2-4. Logic of BCD output
The logic of BCD output is selected by the setting of function F-32.
Function No.
Set value
Contents
F-32
0000
~
1111
0 0
0
0
Digital output logic (Load data, Decimal point data)
POL. logic (Polarity)
Flag logic (ERROR, OVER, Stable, Gross value)
P.C. logic (Print command)
0 : Negative logic
1 : Positive logic
Default is set to [All negative logic].
Содержание CSD-903-EX
Страница 1: ...DIGITAL INDICATOR CSD 903 EX Instruction Manual EN294 1492 H ...
Страница 2: ...IV ...
Страница 91: ...75 ...
Страница 93: ...77 ...
Страница 95: ...79 ...
Страница 97: ...81 ...
Страница 129: ...113 14 4 Connecting method 14 4 1 One to one RS 485 RS 485 S G A B address 00 S G RS 485 host CSD 903 ...
Страница 199: ...183 Inform us about the contents of failure and situation at site in details ...
Страница 204: ...188 Inform us about the contents of failure and situation at site in details ...
Страница 206: ...190 Inform us about the contents of failure and situation at site in details ...
Страница 209: ...193 Inform us about the contents of failure and situation at site in details ...
Страница 224: ... ...
Страница 225: ......