15-1. Calibration by transmission through Modbus interface
Please proceed the setting according to the following flow chart in this calibration.
15-1-1. Calibration mode
Each item of the flow chart corresponds to each procedure of [4-3. Calibration procedure details].
Please refer to [4-3. Calibration procedure details] for the details.
・
While executing the calibration by transmission,
(–RS–) is displayed.
・
When the calibration is compulsorily finished, all the data returns to the condition before the
calibration without executing the registration.
・
Please register the data for the calibration with the weight after confirming the stable condition
without fail.
(1) Change to Calibration mode by
communication
(2) Setting of Scale interval
(3) Setting of Weighing capacity value
(4) Setting of mass of weight
(7) ZERO calibration by numeric input
(10) SPAN calibration by numeric input
(5) Confirmation of stable at ZERO
(6) ZERO calibration by measuring value
(8) Confirmation of stable of SPAN
(9) SPAN calibration by weight
(11) Calibration is finished
In the calibration by numeric input
(Electrical calibration)
In the calibration by actual weight
(Calibration by actual weight)
Содержание CSD-903-EX
Страница 1: ...DIGITAL INDICATOR CSD 903 EX Instruction Manual EN294 1492 H ...
Страница 2: ...IV ...
Страница 91: ...75 ...
Страница 93: ...77 ...
Страница 95: ...79 ...
Страница 97: ...81 ...
Страница 129: ...113 14 4 Connecting method 14 4 1 One to one RS 485 RS 485 S G A B address 00 S G RS 485 host CSD 903 ...
Страница 199: ...183 Inform us about the contents of failure and situation at site in details ...
Страница 204: ...188 Inform us about the contents of failure and situation at site in details ...
Страница 206: ...190 Inform us about the contents of failure and situation at site in details ...
Страница 209: ...193 Inform us about the contents of failure and situation at site in details ...
Страница 224: ... ...
Страница 225: ......