RS-422 check
No.
Correct connecting?
Yes
Connect wires surely with an external
equipment according to the 16-3-11.
No.
The address number is
correctly set.
Yes
Make the change of RS-422/485 correctly.
No.
The communication
protocol is correctly set.
Yes
Set the flow control to “NONE” from “YES”.
No.
Command/data is correctly set.
Yes
Set the address number correctly according
to the 16-3-6.
No.
The host specifies the flow
control.
Yes
Please match the setting of function correctly
to the specification of the host according.
No.
The change of RS-422/485
is properly set.
Yes
Use the correct command/data by referring to
the 16-3-12.
. Inform Minebea about the contents of
failure and situation in details.
Inform us about the contents of
failure and situation at site in details