Scaling of analog output when set value of CF-03 is set to 0 or 1.
CF-71 set the display when the minimum value is output.
CF-72 set the display when the maximum value is output.
Ex) CF-71: set as 1 000
CF-72: set as 5 000
When the display is 5 000, the maximum value is output.
When the display is 1 000, the minimum value is output.
・
The setting of CF-72 must not exceed the weighing capacity set in clause 4.
・
Please set the smaller value in CF-71 than the value set in CF-72.
最大表示
表示
最大表示の+110 %
アナログ出力
約21.6 mA
約11 V
最大値
20 mA
10 V
最小値 4 mA
0 V
0表示
約2.4 mA
約-1 V
最大表示の-10 %
Display
+110 % of the maximum display
The maximum display
+110 % of the maximum display
Display
Approx.2.4 mA
Approx.
–
1 V
-10 % of the maximum display
the maximum display
Approx.21.6 mA
Approx. 11 V
The minimum display
Analog output
5 000
表示
5 500
アナログ出力
約21.6 mA
約11 V
最大値
20 mA
10 V
最小値 4 mA
0 V
1 000
約2.4 mA
約-1 V
最大表示の-10 %
Analog output
Approx.21.6 mA
Approx. 11 V
the maximum display
-10 % of the maximum display
The minimum display
Approx.2.4 mA
Approx.
–
1 V
Display
Содержание CSD-903-EX
Страница 1: ...DIGITAL INDICATOR CSD 903 EX Instruction Manual EN294 1492 H ...
Страница 2: ...IV ...
Страница 91: ...75 ...
Страница 93: ...77 ...
Страница 95: ...79 ...
Страница 97: ...81 ...
Страница 129: ...113 14 4 Connecting method 14 4 1 One to one RS 485 RS 485 S G A B address 00 S G RS 485 host CSD 903 ...
Страница 199: ...183 Inform us about the contents of failure and situation at site in details ...
Страница 204: ...188 Inform us about the contents of failure and situation at site in details ...
Страница 206: ...190 Inform us about the contents of failure and situation at site in details ...
Страница 209: ...193 Inform us about the contents of failure and situation at site in details ...
Страница 224: ... ...
Страница 225: ......