Accutorr 7 Operator’s Manual
D - 5
D.2.2
SpO
2
Alarm Messages
Alarm Message
Clearable?
(Yes/No)
Alarm
Level
Cause
Solution
SpO
2
Sensor Off
(Masimo, Nellcor)
Yes
Low*(
Monito
r
mode)
Off*
(Spot
Check
mode)
The SpO
2
sensor has
become detached
from the patient or
the module, or there
is a fault with the
SpO
2
sensor, or an
unspecified SpO
2
sensor has been
used.
Check the sensor
application site and
the sensor type, and
make sure if the
sensor is damaged.
Reconnect the
sensor or use a new
sensor.
SpO
2
Sensor
Fault
(Masimo)
No
Low
SpO
2
No Sensor
(Masimo, Nellcor)
Yes
Low
SpO
2
Unknown
Sensor (Masimo)
No
Low
SpO
2
Too Much
Light (Masimo)
No
Low
There is too much
light on the SpO
2
sensor.
Move the sensor to
a place with lower
level of ambient
light or cover the
sensor to minimize
the ambient light.
SpO
2
No Pulse
(Nellcor)
No
Low
SpO
2
sensor failed
to obtain pulse
signal.
Move the sensor to
a site with better
perfusion.
SpO
2
Comm Stop
(Masimo, Nellcor)
No
High
An error occurred to
the SpO
2
module, or
there is a problem
with the
communications
between the
module and the
monitor.
Restart the monitor.
If the error remains,
contact Mindray or
your service
personnel.
SpO
2
Comm
Abnormal
(Masimo, Nellcor)
Yes
High
SpO
2
Init Err
Yes
High
SpO
2
Board Fault
(Masimo)
No
Low
There is a problem
with the SpO
2
measurement
board.
Do not use the
module and contact
Mindray or your
service personnel.
Содержание Accutorr 7
Страница 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Страница 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 199: ......
Страница 200: ...P N 046 005276 00 7 0 SW Version 03 10 XX ...