Accutorr 7 Operator’s Manual
12 - 7
■
Exit the scoring screen
■
Manually input a score parameter
12.7
Managing Scorings
12.7.1
Allowing Measured Values to be Edited
The monitor displays the measured parameter values in the scoring screen. By default, if
these measured parameters values are valid, you cannot edit them. If you want to be
able to edit these parameter values, follow this procedure:
1.
Select [
Main
]
→
[
Maintenance>>
]
→
[
User Settings>>
]
→
Enter the required
password
→
[
Manage Scoring
>>]
.
2.
Check the check box of [
Allow Measured Values to be Edited
].
3.
Access the scoring screen and select a desired parameter value to edit.
12.7.2
Importing Scoring
You can import MEWS, NEWS and customized scorings into the monitor. Up to five
scoring protocols can be imported into the monitor.
1.
Connect the USB drive to the USB connector on the monitor.
2.
Select [
Main
]
→
[
Maintenance>>
]
→
[
User Settings>>
]
→
Enter the required
password
→
[
Manage Scoring
>>]
→
[
Import Scoring
>>].
3.
In the [
Import Scoring
] menu, select the scorings to be imported. Then select
[
Import
]. A status message will report completion of the import.
NOTE
•
To load the customized scoring protocols to the USB drive, refer to
Clinical
Scoring Config Instruction for Use
(P/N: 046-007126-00).
12.7.3
Selecting a Default Scoring
The monitor does not provide default scoring. To select a default scoring:
1.
Select [
Main
]
→
[
Maintenance>>]
→
[
User Settings>>
]
→
Enter the required
password
→
[
Manage Scoring>>
]
→
[
Select Default Scoring>>
].
2.
Set the default scoring for the applicable patient category.
After the default scoring is set, when a patient category is changed, the monitor will
automatically use the default scoring.
12.7.4
Deleting a Scoring
To delete the scoring:
Содержание Accutorr 7
Страница 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Страница 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 199: ......
Страница 200: ...P N 046 005276 00 7 0 SW Version 03 10 XX ...