Accutorr 7 Operator’s Manual
9 - 3
■
Rapid blood pressure changes
■
Severe shock or hypothermia that reduces blood flow to the peripheries
■
Obesity, where a thick layer of fat surrounding a limb dampens the oscillations
9.4
NIBP Measurement Mode
There are the following modes of measuring NIBP:
■
Manual: measurement on demand.
■
Auto: continually repeated measurements at set intervals.
■
STAT: continually rapid series of measurements over a five minute period, then
return to the previous mode.
■
Program: continually automatic measurement at set durations and intervals.
To set NIBP measurement mode, select NIBP parameter area to access [
NIBP Setup
]
menu:
■
Select [
Interval
] to select manual mode or auto NIBP measurement interval.
■
Select [
NIBP STAT
] to start a continuous NIBP measurement.
9.5
Measuring
NIBP
9.5.1
Preparing the Patient
In normal use, perform NIBP measurement on a patient who is in the following position:
■
Comfortably seated
■
Legs uncrossed
■
Feet flat on the floor
■
Back, arm and feet supported
■
Middle of the cuff at the level of the right atrium of the heart
.
NOTE
•
It is recommended that the patient relax as much as possible before the NIBP
measurement is performed and that the patient does not talk during
measurement.
•
It is recommended that the patient sit still for 5 min before the first
measurement is taken.
•
The operator should not touch the cuff and tubing during the NIBP
measurement.
Содержание Accutorr 7
Страница 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Страница 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 199: ......
Страница 200: ...P N 046 005276 00 7 0 SW Version 03 10 XX ...