Accutorr 7 Operator’s Manual
A - 13
Effect of interference gases on CO
2
measurements
Gas
Concentration (%)
Quantitive effect*
N
2
O
≤60
±1 mmHg
Hal
≤4
Sev
≤5
Iso
≤5
Enf
≤5
O
2
≤100
Des
≤15
±2 mmHg
*: means an extra error should be added in case of gas interference when CO
2
measurements are performed between 0-40mmHg.
Inaccuracy specifications are affected by the breath rate and I: E change. The EtCO
2
accuracy is within specification for breath rate ≤ 60rpm and I/E ratio ≤1:1, or breath rate
≤30rpm and I/E ratio ≤ 2:1.
Содержание Accutorr 7
Страница 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Страница 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 199: ......
Страница 200: ...P N 046 005276 00 7 0 SW Version 03 10 XX ...