11 - 4
Accutorr 7 Operator’s Manual
◆
For non-intubated patients, place the nasal cannula onto the patient as
shown in the figure below.
4.
Connect the CO
2
gas outlet to the scavenging system with an exhaust tube.
CO
2
can be measured after the start-up is complete.
NOTE
•
If not necessary, do not disconnect the CO
2
adapter from the
equipment after the first installation.This reduces the risk of the CO
2
adapter becoming lost or damaged.
11.7
Changing CO
2
Settings
11.7.1
Changing CO
2
Alarm Settings
To change the CO
2
alarm settings, follow this procedure:
1.
Select the CO
2
numeric area or waveform area to enter the [
CO2 Setup
] menu.
2.
Select [
Alarm Setup>>
].
3.
Set the following alarm properties:
◆
Switch on or switch off the alarms.
◆
Adjust the alarm limits and alarm priority.
11.7.2
Setting the “No CO2 Changes Detected” Alarm Delay
The monitor will alarm if the patient has stopped breathing for longer than the preset
time of the “No CO2 Changes Detected” Delay.
To set the delay time, follow this procedure:
1.
Select the CO
2
numeric area or waveform area to enter the [
CO2 Setup
] menu.
2.
Select [
Delay Time
] and then select the appropriate setting.
Содержание Accutorr 7
Страница 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Страница 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 199: ......
Страница 200: ...P N 046 005276 00 7 0 SW Version 03 10 XX ...