11 - 2
Accutorr 7 Operator’s Manual
11.3
Measurement Limitations
The following factors may influence the measurement accuracy:
■
Leaks or internal venting of sampled gas
■
Mechanical shock
■
Cyclic pressure up to 10 kPa (100 cmH
2
O)
■
Other sources of interference, if any
Measurement accuracy may be affected by the breath rate and inspiration/expiration
ratio (I: E ratio) as follows:
■
EtCO
2
value is within specification for breath rate
≤
60 bpm and I/E ratio
≤
1:1;
■
EtCO
2
value is within specification for breath rate
≤
30 bpm and I/E ratio
≤
2:1.
Measurement accuracy is unspecified for breath rate larger than 60 bpm.
11.4
CO
2
Display
The CO
2
parameter and waveform areas provide FiCO
2
measurement, EtCO
2
measurement, awRR measurement, and a CO
2
waveform.
1.
End tidal CO
2
(EtCO
2
)
2.
Fraction of inspired CO
2
(FiCO
2
)
3.
Airway respiration rate (RR)
4.
CO
2
waveform
5.
Scale
11.5
Automatic CO
2
Module Zeroing
The CO
2
module performs zero calibration automatically when needed.
NOTE
•
The CO
2
module temporally stops measuring during zeroing.
1 2 3
4
5
Содержание Accutorr 7
Страница 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Страница 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 199: ......
Страница 200: ...P N 046 005276 00 7 0 SW Version 03 10 XX ...