Accutorr 7 Operator’s Manual
14 - 1
14
Network
14.1
Network Connection
The monitor can be connected to the central station through the network.
■
The Accutorr 7 transmits waveforms and numerics of parameters (Temp, SpO
2
,
NIBP), related alarms and alarm settings, patient information, and operating mode
to the central station. The waveforms, numerics, alarms, alarm settings, patient
information, and operating mode displayed on central station are consistent with
Accutorr 7.
■
The central station transmits alarm settings, parameter settings, patient
information and operating mode settings (enter or exit the Standby mode) to the
Accutorr 7. The alarm settings, parameter settings, patients’ information and
operating mode settings of the Accutorr 7 are consistent with central station.
The monitor can be connected to the ADT system and the EMR system through the
eGateway.
■
The Accutorr 7 transmits the query command message to the ADT system. When
the ADT system receives the query command message, the ADT system will
transmit the patients’ information to the Accutorr 7.
■
The Accutorr 7 transmits numerics of Temp, SpO
2
, NIBP, related alarms and alarm
settings, patient information, operating mode, and historical data, including
trends and events, to the EMR. When the EMR receives the data, the EMR will send
a success message to the Accutorr 7.
We recommend that you implement the following cybersecurity controls on any
computer that is used to receive data by network communication from the monitors:
■
The computer should be protected by up-to-date antivirus software;
■
The computer should contain access controls to limit its use to authorized users;
Note that if the data transmitted by network is lost, there is also data backup on the
device.
CAUTION
•
Disconnecting from the network may result in data loss, including parameter
waveforms and measurements, alarm events, trends and patient data, or
cause functional failure. In the case of network disconnection, check the
patient and solve the network problem as soon as possible.
Содержание Accutorr 7
Страница 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Страница 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 199: ......
Страница 200: ...P N 046 005276 00 7 0 SW Version 03 10 XX ...