Accutorr 7 Operator’s Manual
D - 11
D.2.6
Power Alarm Messages
Alarm Message
Clearable?
(Yes/No)
Alarm
Level
Cause
Solution
12V Too High
No
High
There is a problem
with the system
power supply.
Restart the monitor.
If the problem still
exists, contact
Mindray or your
service personnel.
12V Too Low
No
High
5V Too High
No
High
5V Too Low
No
High
3.3V Too High
No
High
3.3V Too Low
No
High
Battery Too Low
No
Med
The battery charge
is too low.
Connect the
monitor to an AC
power source and
allow the batteries
to charge
immediately
.
Battery Depleted
No
High
The battery charge
is almost depleted
.
Power Board
Comm Err
No
High
No data from power
module has been
received for 5
seconds.
Restart the monitor.
If the problem still
exists, contact
Mindray or your
service personnel.
RT Clock Need
Reset
No
Low
There is a problem
with the button cell.
Reset the system
time and restart the
monitor. If the
problem still exists,
contact Mindray or
your service
personnel.
PWR interrupted.
Check meas.
State.
Yes
Low
Power supply to the
monitor was
interupted
.
Check the
measurements
when the monitor
restarts. If the
problem still exists,
contact Mindray or
your service
personnel.
Содержание Accutorr 7
Страница 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Страница 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 199: ......
Страница 200: ...P N 046 005276 00 7 0 SW Version 03 10 XX ...