Accutorr 7 Operator’s Manual
Guidance and Declaration — Electromagnetic Immunity
The device is intended for use in the specified electromagnetic environment. The customer or
the user of the device should assure that it is used in such an environment as described
below.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Conduced RF
IEC 61000-4-6
3 Vrms
150kHz to
80MHz
3 Vrms
Portable and mobile RF communications
equipment should be used no closer to
any part of the system, including cables,
than the recommended separation
distance calculated from the equation
appropriate for the frequency of the
transmitter. Recommended separation
distances:
Radiated RF
IEC 61000-4-3
3 V/m 80MHz
to 2.5GHz
3 V/m
Recommended separation distance:
80 M to 800 MHz
800 M to 2.5 GHz
Where, P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
b
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
a
, should be less than the
compliance level in each frequency range
b
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Note 1: At 80 MHz to 800 MHz, the separation distance for the higher frequency range
applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Note 3: The device that intentionally receives RF electromagnetic energy at the exclusion
band (2395.825MHz-2487.645MHz) is exempt from the essential performance requirements,
but remains safe.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the [ME EQUIPMENT or ME
SYSTEM] is used exceeds the applicable RF compliance level above, the [ME EQUIPMENT or
ME SYSTEM] should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the [ME
EQUIPMENT or ME SYSTEM].
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than3V/m.
d
1.2 P
=
d
1.2 P
=
d
2.3 P
=
Содержание Accutorr 7
Страница 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Страница 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 199: ......
Страница 200: ...P N 046 005276 00 7 0 SW Version 03 10 XX ...