Accutorr 7 Operator’s Manual
16 - 7
5.
Open the valve to start flowing CO
2
and make sure that there is sufficient flow
sufficient to vent to atmosphere.
6.
In the [
Calibrate CO2
] menu, enter the vented CO
2
concentration in the [
CO2
]
field.
7.
In the [
Calibrate CO2
] menu, the measured CO
2
concentration is displayed. After
the measured CO
2
concentration becomes stable, select [
Calibrate
] to calibrate
the CO
2
module.
If the calibration is completed successfully, the message [
Calibration Completed!
] is
displayed in the [
Calibrate CO2
] menu. If the calibration fails, the message [
Calibration
Failed!
] is displayed. In this case, attempt to calibrate the CO
2
module again. If the
problem still exists, contact Mindray service personnel.
16.8
Battery Check
Refer to
15.5.2 Checking a Battery
for battery check instructions.
16.9
Calibrating the Touchscreen
Recalibrate the touchscreen whenever the touch interface becomes difficult to
maneuver.
1.
Select [
Main
]
→
[
Maintenance>>
]
→
[
Touchscreen Calibration
]. The
symbol will appear at different positions of the screen.
2.
Select, in turn the central point of the
symbol. After the calibration is
completed, the message [
Screen Calibration Completed!
] is displayed.
3.
Select [
Ok
] to confirm the completion of the calibration.
Monitor
Cylinder
T-shape connector
Tubing
Relief valve
Содержание Accutorr 7
Страница 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Страница 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 199: ......
Страница 200: ...P N 046 005276 00 7 0 SW Version 03 10 XX ...