Alarm conditions are indicated to nurses only when:
1.
The nurse call system is enabled
2.
An alarm that meets your preset requirements occurs, and
3.
The monitor is not in the alarm paused or silence status.
If no setting is selected from [
Alarm Level
] or [
Alarm Category
], no nurse call signal will
be triggered if alarms occur.
WARNING
•
To obtain the nurse call signal, use the nurse call cable (
PN: 009-003116-00
) we
supply. Otherwise the nurse call function will not work and the monitor may
be damaged.
•
Do not rely exclusively on the nurse call system for alarm notification.
Remember that the most reliable alarm notification combines audible and
visual alarm indications with the patient’s clinical condition.
Содержание Accutorr 7
Страница 1: ...Accutorr 7 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 22: ...1 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 30: ...2 8 Accutorr 7 Operator s Manual Trend Screen 2 6 3 9 5 1 8 7 ...
Страница 76: ...6 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 84: ...7 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 86: ...8 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 96: ...9 10 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 102: ...10 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 136: ...15 6 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 164: ...A 14 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 172: ...B 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 180: ...C 8 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 198: ...F 2 Accutorr 7 Operator s Manual This page intentionally left blank ...
Страница 199: ......
Страница 200: ...P N 046 005276 00 7 0 SW Version 03 10 XX ...