MEAS :CONF :TEMP :THER :TYPE (?) <CPD> { B | E | J | K | N | R | S | T }
This command selects the function TC of internal multimeter, sets the type of thermocouple and switches the
multimeter on.
Example: to set the multimeter to thermocouple R measurement mode:
MEAS :CONF :TEMP : THER :TYPE R <cr>
If query is sent, M142 returns the set type of thermocouple { B | E | J | K | N | R | S | T }
MEAS :CONF :TEMP :THER :RJUNction (?) <DNPD>
This command selects the function TC of internal multimeter, sets the temperature of cold junction of selected
thermocouple and switches the multimeter on.
<DNPD>
Represents the temperature expressed in the unit set by the ‘UNIT’ command. In “Technical Data” chapter there
is list of acceptable ranges.
Example: to set the temperature of cold end of thermocouple to 25
C:
MEAS :CONF :TEMP :THER :RJUN 25 <cr>
If query is sent, M142 returns the set temperature using standard exponential format. Example: 20.5
C is
returned as 2.001.
MEAS :CONF :OFF
This command switches the multimeter off.
TEST :RUN <DNPD>
This command starts selected procedure. Number 1 to 10 can be set.
<DNPD>
Represents number of procedure.
Example: to run test procedure 3 :
TEST :RUN 3 <cr>
TEST :RESU ?
M142 returns the result of selected procedure { PASS | FAIL | RUN }.
String „RUN“ is returned if selected procedure isn’t finished.
I/D (instrument identification)
*IDN?
This command returns the identification of the manufacturer, model, serial number and firmware revision.