![Meatest M525 Скачать руководство пользователя страница 114](http://html.mh-extra.com/html/meatest/m525/m525_operation-manual_1758444114.webp)
M-142 Multifunction Calibrator
MEATEST, s.r.o.
114
User Manual v54
voltage range:
1 mV to 200 V
wave form:
square, positive, negative, symmetrical, ramp A, ramp B, triangle,
truncated sin with THD 13,45 %
peak value uncertainty:
0.3 % + 50 uV
displayed values:
peak, effective
min. frequency
0.1 Hz for squarewave, 20 Hz otherwise
Current
summary range DCI:
0
A - 30 A (up to 1500 A with 140-50 Current Coil
*1
)
summary range ACI:
1
A – 30 A (up to 1500 A with 140-50 Current Coil
*1
)
internal ranges:
200
A, 2 mA, 20 mA, 200 mA, 2 A, 30 A
resolution:
61/2 digit
frequency resolution:
6 digit, minimum step 0.001Hz
frequency uncertainty:
0.005 %
frequency range:
20 Hz to 5 kHz
(up to 200 µA)
20 Hz to 10 kHz (0.2 – 200 mA)
20 Hz to 1 kHz
(0.2 – 20 A)
40 Hz to 500 Hz (20 – 30 A)
DCI uncertainty
Range
% value + μA
Min. compliance voltage [V]
0.0000
A – 200.0000
A
0.05 + 0.02
3
0.200000 mA – 2.000000 mA
0.02 + 0.1
3
2.00000 mA – 20.00000 mA
0.01 + 0.6
3
20.0000 mA – 200.0000 mA
0.01 + 6
3
0.200000 A – 2.000000 A
0.015 + 100
3
2.00000 A – 30.00000 A
*2, *3
0.02 + 2 000
1.5
ACI uncertainty
Range
% value + μA Min. compliance
voltage [Vrms]
% value + μA
Min. compliance
voltage [Vrms]
% value + μA Min. compliance
voltage [Vrms]
20 Hz - 1 kHz
1 kHz – 5 kHz
5 kHz – 10 kHz
1.0000
A – 200.0000
A
0.15 + 0.02
3
0.30 +0.22
3
- -
- -
0.200000 mA – 2.000000 mA
0.07 + 0.2
3
0.20 + 1
3
0.50 + 1.4
2
2.00000 mA – 20.00000 mA
0.05 + 1
3
0.20 + 10
3
0.50 + 14
2
20.0000 mA – 200.0000 mA
0.05 + 10
3
0.20 + 100
3
0.50 + 140
2
0.200000 A – 2.00000 A
0.05 + 100
3
- -
- -
- -
- -
2.00000 A – 30.00000 A
*2, *3
0.1 + 6000
1.5
- -
- -
- -
- -
*1
140-50 Current Coil increases output uncertainty by additional 0.3% of value. Available multiplication factors: x25 and x50.
Recommended frequency range is 50-60 Hz, combination of high frequency and amplitude might overload current output.
*2
10 to 20 A: Continuous output time is limited to 60 s.
*3
20 to 30 A: Continuous output time is limited to 30 s. Frequency is limited to 40 – 500 Hz. Minimum compliance voltage 0.5 Vrms. Add
additional uncertainty 900*(I-20) μA, where I is set output current in [A].
Auxiliary parameters
Range
200 uA
2 mA
20 mA
200 mA
2 A
30 A
max. inductive load
*1
1H
100 mH
100 mH
10 mH
1 mH
500 uH
THD
*2
0,2%
0,2%
0,2%
0,2%
0,2%
0,3%
*3
*1
DC and AC 50/60 Hz, voltage on the load < 2Vrms
*2
Parameter includes non-linear distortion and non-harmonic noise
*3
Distortion in range 20 to 30 A is max. 0.5%
Function Shape
current range:
100 uA to 2 A
wave form:
square, positive, negative, symmetrical, ramp A, ramp B, triangle,
truncated sin with THD 13,45 %
peak value uncertainty:
0.3 % + 500 nA
displayed values:
peak, effective
min. frequency
0.1 Hz for squarewave, 20 Hz otherwise
Resistance
summary range:
0
to 1000 M