MEATEST, s.r.o.
M-142 Multifunction Calibrator
User Manual v54
53
Press “EXIT” display button to terminate the calibration. After the button is pressed, the calibration date is saved
internally and the calibrator returns to the state it was in before the calibration has been started.
Calibration points
Each function of the calibrator has assigned fixed calibration points which have to be set during the calibration.
For VOLTAGE DC, VOLTAGE AC, CURRENT DC, CURRENT AC, POWER AC, POWER DC, F functions,
the signal value is set using the keyboard. For R-C functions and partially also during the calibration of the
multimeter, calibration data of the range resistances must be entered. T function does not require any calibration,
as the output voltage or resistance is based on arithmetic interpolation using standard tables of temperature
sensor values.
The calibrator needs no calibration of the following parameters:
•
frequency
•
phase relationship (power factor) of output voltage and frequency in AC power/energy generation
mode
•
DC and AC power (except DC and AC current in this mode)