MEATEST, s.r.o.
M-142 Multifunction Calibrator
User Manual v54
81
Tables of limits
20 V DC Basic range with linearity test
Function
Range
Value (V)
Frequency (Hz) Deviation
allowed(
μ
V)
V-DC
20.0 V
2.0
DC
70
V-DC
20.0 V
4.0
DC
90
V-DC
20.0 V
6.0
DC
110
V-DC
20.0 V
8.0
DC
130
V-DC
20.0 V
10.0
DC
150
V-DC
20.0 V
12.0
DC
170
V-DC
20.0 V
14.0
DC
190
V-DC
20.0 V
16.0
DC
210
V-DC
20.0 V
18.0
DC
230
V-DC
20.0 V
19.0
DC
240
V-DC
20.0 V
-2.0
DC
70
V-DC
20.0 V
-4.0
DC
90
V-DC
20.0 V
-6.0
DC
110
V-DC
20.0 V
-8.0
DC
130
V-DC
20.0 V
-10.0
DC
150
V-DC
20.0 V
-12.0
DC
170
V-DC
20.0 V
-14.0
DC
190
V-DC
20.0 V
-16.0
DC
210
V-DC
20.0 V
-18.0
DC
230
V-DC
20.0 V
-19.0
DC
240
Table I
DC voltage test
Function
Range
Value (V)
Frequency (Hz) Deviation
allowed(V)
V-DC
2.0 V
1.9
DC
33
μ
V-DC
2.0 V
-1.9
DC
33
μ
V-DC
240.0 V
190.0
DC
3.335 m
V-DC
240.0 V
240.0
DC
4.100 m
V-DC
240.0 V
-190.0
DC
3.335 m
V-DC
240.0 V
-240.0
DC
4.100 m
V-DC
1000.0 V
1000.0
DC
55 m
V-DC
1000.0 V
-1000.0
DC
55 m
Table II
20 V AC Basic range with linearity test
Function
Range
Value (V)
Frequency (Hz) Deviation
allowed(%value)
V-AC
20.0 V
2.0
1000
1.36 m
V-AC
20.0 V
4.0
1000
1.72 m
V-AC
20.0 V
6.0
1000
2.08 m
V-AC
20.0 V
8.0
1000
2.44 m
V-AC
20.0 V
10.0
1000
2.80 m
V-AC
20.0 V
12.0
1000
3.16 m
V-AC
20.0 V
14.0
1000
3.52 m
V-AC
20.0 V
16.0
1000
3.88 m
V-AC
20.0 V
18.0
1000
4.24 m
V-AC
20.0 V
19.0
1000
4.42 m
Table III