M-142 Multifunction Calibrator
MEATEST, s.r.o.
68
User Manual v54
No
error
label
description
troubleshooting
01 Overload 2V !
2V range overloaded
Output current is too high. Increase load resistance.
02 Overload 20V !
20 V range overloaded
Output current is too high. Increase load resistance.
03 Overload 200V !
200, 1000 V ranges
overloaded
Output current is too high. Increase load resistance.
04 Overload I output ! Current output overloaded
Voltage on the load is too high. Decrease load resistance.
05 High temperature ! Too high internal
temperature
Output stages are overloaded. Do not use ranges 200V, 1000V or
20 A for at least 10 minutes. Check if the ventilation holes are
free.
06 Overload RC !
RC simulator overloaded
Test current is too high. Use lower range of tested Ohmmeter.
07
FBK error !
Internal error
Turn off the calibrator and turn on it again.
08
OUTPUT must be
in OFF state !
Cable adapter tried to be
exchanged while output
terminals was ON
Switch off output terminals with button OUTPUT, change the
adapter, and switch output terminals on.
10
Interface error !
GPIB communication error Wrong data format on GPIB.
11 Bad command !
Bad command of GPIB
Not known command on GPIB.
12 Bad
communication !
GPIB communication error Listener not connected to the GPIB. Check correct connection of
GPIB cable.
13 Over range !
Overcrossing the range via
GPIB
Value out of range was set via GPIB. Set correct value.
20 Bad calib. code !
Bad calibration code
Wrong calibration code was entered, calibration cannot start.
Enter correct calibration code.
21 Time warm up !
Attempt to start calibration
before warm up
Attempt to start calibration before 60 minutes warm up period.
Let the calibrator turned on for at least 60 minutes.
24 Cable adapter must
be off !
Adapter isn’t allowed for
autocalibration
Use another cable adapter or perform autocalibration procedure
without cable adapter.
25 Use cable adapter ! Attempt to start calibration
without cable adapter.
Calibration of resistance ranges can be performed with adapter
Option 70. Calibration of internal multimeter can be performed
with cable adapters Option 40 and Option 60.
30 Internal RxD
timeout !
Internal error
Internal error of the calibrator. Turn the calibrator off and after 5
s turn on. If the error will appear again, contact manufacturer.
31 Internal
communication !
Internal error
Internal error of the calibrator. Turn the calibrator off and after 5
s turn on. If the error will appear again, contact manufacturer.
37 Calibrator is not
ready !
Internal error
Internal error of the calibrator. Turn the calibrator off and after 5
s turn on. If the error will appear again, contact manufacturer.
40 Value too large !
Maximum value is out of
limit
Attempt to set value over possible range. Set correct value.
41 Value too small !
Minimum value is out of
limit
Attempt to set value under possible range. Set correct value.
42 Deviation too large
!
Deviation is too high
Set deviation is out of limit –30% to +30%. Set correct value.
44 /- !
Change of polarity is not
allowed
Attempt to change polarity, where it is not allowed.. Concerned
modes F, P-E, R-C, ACV, ACI.
45 Unable – polarity ! Negative polarity is not
allowed.
Attempt to set negative polarity, where it is not allowed..
Concerned modes F, P-E, R-C, ACV, ACI.
46 Unable DC/AC !
DC/AC conversion is not
possible
Attempt to change parameter AC/DC where it is nonsense or
where it is not allowed.
47
Current timeout
!
Time limit for current over
10 A exceeded
Long-term loading current terminals with output current over A.
48
Not allowed on
AUX output !
Function cannot be used on
output AUXILIARY
Do not use this setting in combination with connected cable
adapter.
Tab. 16 Error messages