MEATEST, s.r.o.
M-142 Multifunction Calibrator
User Manual v54
57
RESISTANCE function
nominal value [
]
set limits [
]
range [
]
0
0.002
10
10
0.002
10
33
0.002
10 – 33
33
0.002
33 – 100
100
0.005
33 – 100
100
0.005
100 – 330
330
0.012
100 – 330
330
0.012
330 – 1 k
1 k
0.025
330 – 1 k
1 k
0.025
1 k – 3k3
3k3
0.100
1 k – 3k3
3k3
0.100
3k3 – 10 k
10 k
0.25
3k3 – 10 k
10 k
0.25
10 k – 33 k
33 k
1
10 k – 33 k
33 k
1
33 k – 100 k
100 k
2.5
33 k – 100 k
100 k
2.5
100 k – 330 k
330 k
10
100 k – 330 k
330 k
10
330 k – 1 M
1 M
50
330 k – 1 M
1 M
50
1 M – 3M3
3M3
400
1 M – 3M3
3M3
400
3M3 – 10 M
10 M
2 k
3M3 – 10 M
10 M
2 k
10 M – 33 M
33 M
15 k
10 M – 33 M
33 M
15 k
33 M – 100 M
100 M
190 M
50 k
100 k
33 M – 100 M
100 M -330 M
1 G
1 M
330 M – 1 G
22 k
0.4
linearity
22 k
0.4
linearity
Tab.12 Resistance
Calibration in resistance range of calibration points from 0 Ω to 100 kΩ must be done using Option 70 cable
adapter (four-terminal connection). Four-wire output resistance is defined at the test leads of the cable adapter.
Calibration in resistance range of calibration points from 100 kΩ to 1 GΩ must be done without Option 70 cable
adapter application (two-terminal connection). Use front panel terminals Hi-Lo for connection of standard
ohmmeter.