M-142 Multifunction Calibrator
MEATEST, s.r.o.
46
User Manual v54
Note
If neither the calibrator’s output, nor the meter’s inputs are grounded, signal/noise ratio can arise at the
calibrator’s output.
4.
Temp.scale .... xx ITS90/PTS68
This parameter allows the temperature scale for platinum resistance temperature sensors to be selected. Pressing
the display buttons allows to switch between ITS90 and PTS68 temperature scales. ITS90 is set by the
manufacturer.
5.
Temp.unit .... xx
o
C/K
This parameter allows the temperature unit for simulation of temperature sensors to be selected. Pressing the
display buttons allows to switch between °C and K. °C is set by the manufacturer.
6.
Phase.unit .... xx
o
/cos
This parameter allows the unit of phase shift between the voltage and current output to be selected in the
power/energy generation mode. Pressing the display buttons allows to switch between ° and COS (degrees and
power factor).
7.
Meter average .... xx
UP/DOWN
Displays the number of measurements taken before the average value of build-in multimeter is displayed
(integration constant). UP, DOWN display buttons can be used to select a value in the range of 1 to 20. The
higher the value, the longer the calibrator needs to measure a value, but displayed value is more stable. If 20 is
set, one measurement takes approximately 2.5 s.
8.
Interface .... xx
GPIB/RS232
Displays the type of interface used to control the calibrator from a PC. By pressing GPIB or RS232 buttons, the
respective type can be selected. The calibrator can be remotely controlled only using the selected interface.
9.
GPIB address .... xx UP/DOWN
Displays the calibrator’s address at the GPIB bus. UP, DOWN display buttons can be used to select any valid
GBIP address in the range of 00 to 30. The address 02 is set by the manufacturer.
10.
RS232 baud rate .... xx
UP/DOWN
Indicates the communication speed of RS232 bus. UP/DOWN display buttons can be used to select 150, 300,
600, 1200, 2400, 4800, 9600, 19200. Perfect communication with the PC requires equal values set at the PC and
the calibrator.
11.
Handshake .... xx
OFF/Xon-Xoff
Indicates the communication handshake. Display buttons can be used to select OFF or Xon/Xoff. Perfect
communication with the PC requires equal values set at the PC and the calibrator.
12.
Keyb.beep .... xx
ON/OFF