M-142 Multifunction Calibrator
MEATEST, s.r.o.
62
User Manual v54
11.
Capacitance ranges calibration
Capacitance ranges calibration is performed by both calibration methods, i.e. setting the checked value
and measuring this value with the standard instrument, as well as setting the nominal value of the
calibrator’s measurement range on the external multimeter by adjusting the calibrator’s output. If the
entry of measured value to the calibrator is required, VALUE METER is displayed at the respective
calibration point. The value on the calibrator’s display has the same format as the nominal value of the
respective calibration point. If, on the other hand, the user is requested to adjust the calibrator’s output
to set the nominal value of the calibrator’s measurement range on the external multimeter, VALUE is
displayed at the respective calibration point. The value on the calibrator’s display has general format.
a)
Select CAPACITANCE from the calibration menu. Select GND U OFF and GND I OFF from
the setup menu.
b)
Adjust the values of shorted and open terminals of the external RLC meter, set the frequency
used during the measurement (1000 Hz for 1 nF to 10 nF range, 100 Hz for 10 nF to 100 uF
ranges). Connect Hi, Hu terminals to calibrator’s Lo output and Li, Lu terminals to calibrator’s
Hi output.
c)
Measure the capacitance of internal range capacitors using the RLC meter and write the values
into the calibrator. The procedure is the same as when calibrating the resistance ranges.
d)
Perform the adjustment for all calibration points as usual. Press WRITE display button to
confirm correct values.