![Meatest M525 Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/meatest/m525/m525_operation-manual_1758444013.webp)
MEATEST, s.r.o.
M-142 Multifunction Calibrator
User Manual v54
13
•
value of R0 resistance and the type of resistance temperature sensor
•
cold junction temperature of TC sensors and the selected type of TC sensor
•
value of amplitude and shape type for frequency function
f)
Information section
The information section located in the right part of the display displays additional information related to
the selected function:
•
symbol of connected or disconnected output terminals.
At the same time, a LED located above the OUTPUT button is lit.
•
information about remote/local control of the calibrator. If the calibrator is controlled remotely,
REM is displayed. If the calibrator is controlled locally using the keyboard, LOCAL is displayed.
•
information about the use of 50-turn coil (COIL x50) at the current output of the calibrator, if this
feature is turned on using the SETUP menu.
•
information about the type of connected cable adapter, if used
•
information about the grounding method of output terminals: GND I, GND U as set up using the
setup menu.
g)
Information about the uncertainty of the output signal
This section displays the maximum error of the main value of the output signal. The value is calculated
using the main specification listed in the User’s Manual and it is displayed in %.
2.
INPUT section
This section displays the values measured by the multimeter. The section includes following data:
a)
Main value of measured signal
This line displays the measured value and the unit of measurement. If the input signals exceed the
permitted range, OVERFLOW is displayed.
b)
Designation of selected function of the multimeter
Symbolic display of selected function of the multimeter: V DC, mA DC, mV DC, R 4W, Freq, T TC, T
RTD, ACAL.
3.
Display buttons section
This line displays the symbolic descriptions which define the meaning of four related display buttons. The
respective meanings are as follows:
symbol
button function
note
x 10
increase set value 10 x
: 10
decrease set value 10 x
Shape
selection of signal shape
only for U, I, F functions
+/-
reversed polarity of output voltage and current
only for DC U, DC I functions
EXIT
move up one level
only for F, P-E functions
Calib.
enter the calibration menu
SETUP
enter the setup menu
TC type
selection of thermocouple sensor type
only for T function
RTD type
selection of resistance temperature sensor type
only for T function
f
enter the frequency of the signal
only for U, I function
MODE
select the unit of measurement
only for AC P-E function
Tab. 4
Display buttons description