Setup menu
The multifunction calibrator allows many other, less frequently used parameters to be set. Setup menu is used to
set these parameters. Setup menu is opened by pressing SETUP display button. If output terminals are
connected, they will be disconnected and the following display appears:
Use
or
cursor button or the knob of the potentiometer to browse the menu options. Active option is always
inverted and when changed, the descriptions of display buttons change as well. Display buttons show how the
respective parameter can be set. Each parameter can be changed after the knob of the potentiometer is pressed.
Press EXIT display button twice to save the parameters when the setting is completed. New settings are retained
when the calibrator is switched off. Setup menu offers the following options:
1.
Current Coil .... xx x25/x50/OFF
This parameter can be set on when 50-turn current coil is going to be used for clamp ammeter calibration. The
coil multiplies the output current either 25 times or 50 times depending on set paarmeter. OFF is set by the
manufacturer.
2.
GND U .... xx
ON/OFF
This parameter connects Lo terminal to GND. In practice this means that Lo terminal is grounded. By pressing
the display buttons, the terminal can be grounded or ungrounded. ON is set by the manufacturer, output terminal
is grounded.
3.
GND I .... xx
ON/OFF
This parameter connects -I to GND. In practice this means that -I terminal is grounded. By pressing the display
buttons, the terminal can be grounded or ungrounded. OFF is set by the manufacturer, output terminals are not
grounded.
It is recommended to ground only the voltage channel GND U ON, GND I OFF, for all ranges except the
generation of power or energy. If the meter to be calibrated has Lo terminal grounded, it is recommended to
unground both outputs of the calibrator, GND U OFF, GND I OFF to exclude ground loops.