MEATEST, s.r.o.
M-142 Multifunction Calibrator
User Manual v54
35
*
main data of temperature in
o
C or K
*
sensor type
thermocouples:
K, N, R, S, B, J, T, E
resistance temperature sensors:
Pt 1.385, Pt 1.392, Ni
* resistance at 0
o
C labeled R0 (resistance temperature sensors only)
* cold junction temperature of thermocouple sensors labeled RJ
*
set value of relate deviation in %, labeled
T = xxxx.x °C (K)
the information section shows:
*
temperature scale type
*
uncertainty of simulated temperature value of selected temperature sensor type
•
Set the main value of temperature using numeric keyboard, cursor buttons or potentiometer. Output
terminals are disconnected, the information section of the display shows
the symbol which shows
that output terminals are disconnected.
•
Connect the object to be calibrated to Hi - Lo terminals.
•
Press OUTPUT button.
•
Red LED is lit above the OUTPUT terminal to indicate that the output signal is connected to output
terminals. The display shows the symbol of connected output terminals.
Note
•
Load of output terminals is limited similarly to corresponding voltage or current ranges.
•
Output signals provided at Hi - Lo terminals and AUXILIARY connector are short-circuit proof.
Switching between temperature sensor types
•
Keep pressing “TC type” or “RTD type” display button to select desired sensor type.
•
If resistance temperature sensors are selected, each press of the button selects Pt1.385, Pt1.392 or Ni
resistance thermometer. The display shows current setting as Pt385 / Pt392 / Ni.
•
If thermocouples are selected, each press of the button selects K, N, R, S, B, J, T, E types. The display
shows current setting as TC TYPE x, where x is the type of the thermocouple