M-142 Multifunction Calibrator
MEATEST, s.r.o.
82
User Manual v54
AC voltage test
Function
Range
Value (V)
Frequency (Hz) Deviation
allowed(V)
V-AC
20 mV
0.019
1000
68
μ
V-AC
200 mV
0.19
1000
270
μ
V-AC
2.0 V
1.9
1000
442
μ
V-AC
20.0 V
19.0
50
4.42 m
V-AC
20.0 V
19.0
120
4.42 m
V-AC
20.0 V
19.0
10000
4.42 m
V-AC
20.0 V
19.0
20000
15.5 m
V-AC
20.0 V
19.0
50000
15.5 m
V-AC
240.0 V
190.0
1000
44.2 m
V-AC
1000 V
750.0
120
500.0 m
Table IV
200 mA DC Basic range with linearity test
Function
Range
Value (A)
Frequency (Hz) Deviation
allowed(
μ
A)
A-DC
200.0 mA
0.02
DC
8
A-DC
200.0 mA
0.04
DC
10
A-DC
200.0 mA
0.06
DC
12
A-DC
200.0 mA
0.08
DC
14
A-DC
200.0 mA
0.10
DC
16
A-DC
200.0 mA
0.12
DC
18
A-DC
200.0 mA
0.14
DC
20
A-DC
200.0 mA
0.16
DC
22
A-DC
200.0 mA
0.18
DC
24
A-DC
200.0 mA
0.19
DC
25
A-DC
200.0 mA
-0.02
DC
8
A-DC
200.0 mA
-0.04
DC
10
A-DC
200.0 mA
-0.06
DC
12
A-DC
200.0 mA
-0.08
DC
14
A-DC
200.0 mA
-0.10
DC
16
A-DC
200.0 mA
-0.12
DC
18
A-DC
200.0 mA
-0.14
DC
20
A-DC
200.0 mA
-0.16
DC
22
A-DC
200.0 mA
-0.18
DC
24
A-DC
200.0 mA
-0.19
DC
25
Table V
DC current test
Function
Range
Value (A)
Frequency (Hz) Deviation
allowed(A)
A-DC
200.0 uA
0.00019
DC
0.115
μ
A-DC
200.0 uA
-0.00019
DC
0.115
μ
A-DC
2.0 mA
0.0019
DC
0.430
μ
A-DC
2.0 mA
-0.0019
DC
0.430
μ
A-DC
20.0 mA
0.019
DC
2.5
μ
A-DC
20.0 mA
-0.019
DC
2.5
μ
Table VI