M-142 Multifunction Calibrator
MEATEST, s.r.o.
80
User Manual v54
30.
Check harmonic distortion of output signal. It should not exceed 0.05%.
If calibrator is out of limits is in some points of this test, appropriate function and range should be recalibrated. It
is not necessary to recalibrate all functions, but only this one, which does not meet specification. See chapter
Calibration mode, where recalibration procedure is described.