MEATEST, s.r.o.
M-142 Multifunction Calibrator
User Manual v54
115
resolution :
41/2 digit
Resistance uncertainty
Range
% of value *
1
*
3
Resolution
[
]
Applicable test current
range
Max. voltage/current
on lower range limit *
2
Max. voltage/current on
higher range limit *
2
0
- 10
0.03 + 5 m
0.01
400 uA - 100 mA
- / 100mA
1V / 100mA
10
- 33
0.015 + 5 m
0.01
400
A - 100 mA
1V / 100mA
3.3V / 100mA
33
- 100
0.010 + 5 m
0.01
400
A – 40 mA
3V / 100 mA
3.3V / 100mA
100
- 330
0.010 + 5 m
0.01
400
A – 40 mA
3V / 30 mA
10V / 30mA
330
- 1 k
0.010
0.1
400
A – 11 mA
3.3V / 10mA
10V / 10mA
1 k
- 3.3 k
0.010
0.1
100
A – 6 mA
12V / 12mA
20V / 6mA
3.3 k
- 10 k
0.010
1
20
A – 2 mA
12V / 3.5mA
20V / 2mA
10 k
- 33 k
0.010
1
4
A – 600
A
12V / 1.2mA
20V / 600uA
33 k
- 100 k
0.010
10
1
A – 200
A
12V / 360uA
20V / 200uA
100 k
- 330 k
0.010
10
1
A – 60
A
12V / 120uA
20V / 60uA
330 k
-1 M
0.010
100
0.2
A – 20
A
12V / 36uA
20V / 20uA
1 M
- 3.3 M
0.020
100
40 nA – 6
A
12V / 12uA
20V / 6uA
3.3 M
- 10M
0.050
1.000
10 nA – 2
A
12V / 3.6uA
20V / 2uA
10 M
- 33 M
0.1
1.000
10 nA – 600 nA
12V / 1.2uA
20V / 0.6uA
33 M
- 100M
0.2
10.000
10 nA – 180 nA
12V / 0.36uA
20V / 0.18uA
100 M
- 1000 M
0.5
10.000
4 nA – 20 nA
12V / 0.1uA
20V / 0.02uA
*1
In range 0
to 200 k
specification is valid for four-terminal connection using Option 70 Cable adapter. For two-terminal connection
from front panel terminal add 20 mOhm.
*2
Voltage is expressed in V pk-pk.
*3
In range 200 k
to 1000 M
specification is valid for two-terminal connection directly at front panel terminals Hi-Lo.
Capacitance
summary range:
0.7 nF to 100
F
resolution :
41/2 digit
Capacitance uncertainty
Range
% of value
Max. frequency [Hz]
Max. test voltage [V]
700 pF – 1 nF
0.5 + 15 pF
1000
2 – 5.5
1 nF – 3.3 nF
0.5 + 5 pF
1000
2 – 5.5
3.3 nF - 10 nF
0.5
1000
2 – 5.5
10 nF – 33 nF
0.5
1000
2 – 5.5
33 nF – 100 nF
0.5
500
2 – 5.5
100 nF – 330 nF
1
300
2 – 5.5
330 nF - 1
F
1
300
2 – 5.5
1
F – 3.3
F
1.5
300
2 – 5.5
3.3
F – 10
F
1.5
300
2 – 5.5
10
F - 100
F
2.0
300
2 – 5.5
AC and DC power/energy
summary voltage range:
0.2 V to 240 V
current capability of voltage output:
depends on the voltage range
summary current range:
2 mA to 20 A
maximal voltage on current output:
depends on the current range
electric power range:
0.0004 to 4.8 kVA
phase angle ( PF range):
-180° to +180° (-1 to +1 in PF)
phase angle ( PF ) resolution:
0.1° (0.001 in PF)
time period range:
1.1 s to 1999 s
time period accuracy:
0.01% + 0.1 s
time setting resolution:
0.1s
frequency range:
DC, 40 Hz to 400 Hz
frequency resolution: