7 Synchronized operation option
96
DOEF series
Matsusada Precision
7-1-2 Setting Synchronous Trigger
In order to use the synchronous trigger function, synchronous master and synchronous slave units
need to be set.
Turning the output on and off is operated by the synchronous master unit. Therefore, be sure to set
up one synchronous master unit in the system with the synchronized operation.
[Setting synchronous master and slave units]
There are setting items for synchronous trigger off, master unit, and slave unit on Main Set 2.
In order to synchronize the operation, be sure to perform this setting first.
Press the SET/SAVE switch to enter the setting state, use the downward "
▼
" movement switch to
select "MORE", and press the ENTER switch to switch from Main Set 1 to Main Set 2.
Use the downward "
▼
" movement switch to select "Sync Trig" and set OFF/ MAS (synchronous
master unit)/ SLV (synchronous slave unit) by turning the DIAL.
Then, press the SET/SAVE switch to save the setting.
Synchronous master unit
Only one unit set in the synchronous system.
Output on/off and sequence run/stop are
operated.
Synchronous slave unit 1
Output on/off operation is not
allowed.
Setting synchronous trigger
OFF: Not used
MAS: Synchronous master unit
SLV: Synchronous slave unit
Fig. Main Set 2
Matsusada
Main Set
2
LOC
DC
Ext Signal
Soft Start
Soft Stop
SYSTEM
Response
Sense
Sync Trig
Sync Clk
CV
AC
OFF
FAST
OFF
DEN
Vdc +10.003 V
Vrms 14.138 V
Irms 14.146 A
Idc +9.997 A
Содержание DOEF Series
Страница 1: ...Instruction Manual F RA 001 3R2 MODEL DOEF series B N 243 9 024 243 9 024 Rev 0 4 ...
Страница 15: ...2 External Appearance 7 DOEF series Matsusada Precision 2 3 Dimensions 2 3 1 400W Model ...
Страница 16: ...2 External Appearance 8 DOEF series Matsusada Precision 2 3 2 600W Model ...
Страница 17: ...2 External Appearance 9 DOEF series Matsusada Precision 2 3 3 1200W Model ...
Страница 121: ......
Страница 123: ......
Страница 124: ......