148
Mahr GmbH • Digimar 817 CLM
Height measuring instrument 817 CLM
13
Technical data
Measuring range
350 mm
600 mm
1000 mm
14“
24“
40“
Application range
respectively ca. 170 mm / 7“
Measuring error
1.8 + L/600 (L in mm)
(at 20 °C, base plate according to DIN 876 /0,
styli 6.0 mm)
Perpendicularity error
< 5 µm
< 6 µm
< 10 µm
(Base plate according to DIN 876 /0 and only in
conjunction with a corrected electronic measuring
system in the X-axis, with an incremental probe
Perpendicularity error mechanic
15 µm
20 µm
30 µm
Repeata/- 2
on a plane: 0.5 µm
in a bore: 1 µm
Measuring force, probe 6.0 mm
1 N +/- 0.2 N
Contacting speeds
5, 8, 11, 15, 20 mm/sec. max. 40 mm/sec.
Max. permissible manual positioning speed of the
600 mm/s
measuring carriage
Drive mechanism
motorizied
3-point air cushion
ca. 9 µm
Compressed air supply
integrated compressor
Interchangeable probes
see accessories
Vertical measuring system of the column
incremental measuring system
Working- / operating temperature
10 °C ... 40 °C (50°F... 104 °F)
Storage temperature
-10 °C...60 °C (14°F... 140 °F)
Permissible relative humidity (operating )
max. 65% (non-condensing)
Permissible relative humidity (storage)
max. 65 % (non-condensing)
Weight ca.
25 kg (55.16 lbs) 30 kg (66.14 lbs)
35 kg (77.16 lbs)
Thermometer / Sensor error limits
+/- 0.5 °C (+/- 31.1°F)
Содержание Digimar 817 CLM
Страница 8: ...8 Mahr GmbH Digimar 817 CLM 1 2 Unpacking 1 2 N I C K L M O P Q...
Страница 10: ...10 Mahr GmbH Digimar 817 CLM 5 350 mm 25 kg 600 mm 30 kg 1000 mm 35 kg...
Страница 11: ...11 Mahr GmbH Digimar 817 CLM 6 45 7 8 90...
Страница 12: ...12 Mahr GmbH Digimar 817 CLM 9 11 G H 10 12 13 M...
Страница 13: ...13 Mahr GmbH Digimar 817 CLM B 3x L M 19 17 18 P D E K N 14 15 16...
Страница 14: ...14 Mahr GmbH Digimar 817 CLM 100 240 V 50 60 Hz I 12 h 21 22 Euro US UK 20 23...
Страница 152: ......