Set the zero point on the base plate
Creating a measuring program
First a program must have at least 5 program steps.
The first measuring point must be on the base plate and the last value should be as close as possible to the
maximum measuring length of the height measuring instrument!
Check with a gage block.
e.g. 0.0 mm ; 20.0 mm ; 70.0 mm ; 110.0 mm ; 180.0 mm ; 240.0 mm ; 350.0 mm
for height measuring instrument with a 350 mm measuring length. Height measuring instruments with a larger
measuring range, the measuring points must be set accordingly.
0
Delete all
Contact from above – contact all values in the
measuring series
0.00 mm -> 20.00 mm -> 70.00 mm etc.
Press the Program key
CE
PROG
1. <-_______
2. Learn programming (teach-in)
3. Create new program
Save program
PRG1
09:57
07/Feb/07
755
TEST1
10:59
07/Feb/07
587
TEST5
10:18
04/Sep/06
587
New file name
CORRECT1
Standard tolerances:
_____0.010______
Return to
„Measuring menu“
CE
Delete all
CE