139
Mahr GmbH • Digimar 817 CLM
Press the ON/OFF key to confirm
and
ON
OFF
exit the menu with the „CE key“
CE
Procedure
Mount the incremental probe (P1514 H) into the holder
for perpendicularity error measurement.
5. Enter password
6. Correction tables
7. Clear functions menu
4. MK height correction
5. Print correction tab
6. Perpend. correction
The maximum measuring path depends upon the
height of the height measuring instrument, enter the
height (Note: ca. 1.0 mm shorter).
The height measuring instrument sets the zero point
automatically on the base plate, and travels the
measuring path, the correction factor will be
automatically be accepted.
To check this function, select the „perpendicularity key“
and manually move the probe upwards. In the display
will appear the actual perpendicularity error of the
column, the graph shows the maximum and minimum
values.
Password allocation and printing see correction
factors in the Z-axis!
Correction value
Active customer correction