39
Mahr GmbH • Digimar 817 CLM
4.1.4 Deviations
Probes that have a large deflection (very long or very thin probes) can
have a probe deviation of > 2.2 µm
In the display is the determined probe constant and the deviation
between the 2 conducted measurement.
The following function keys appear:
Conduct calibration once more, the average (mean) value will be
calculated from the previous determined probe constant and the newly
measured constant.
Accept the deviation and value.
Probe dia. ? will appear in the display
Deviation is not accepted, the old value will be maintained .
Abort measurement, all previous probe constants will be maintained.
Description / Sequence
Symbols / Pictures