100
Mahr GmbH • Digimar 817 CLM
The height measuring instrument can save a
measuring process and create a measuring program
from the saved data.
Measuring programs can be created in order to
automate recurring measuring processes. Since any
characteristic can be detailed in an inspection plan, the
measured values can be saved so that they can be
read at any time and evaluated statistically.
A maximum of 40 measuring programs can be saved
permanently in the mass memory, up to 8000
measuring values can be saved in the measuring value
mass memory. All measuring values recorded during a
program run are saved under the name of the program.
The file names may comprise of a maximum of 15
alphanumerical characters.
Fundamental is:
-
The probe is correctly calibrated
-
A suitable die-down time has been set
-
All stored measurements be deleted from the
memory
-
Measure the complete workpiece (also computation
functions such as distance, symmetry...)
7.
Measuring program
1. <-_______
2. Learn programming (teach - in)
3. Create new program
4. Change existing program
5. Print out current program
6. UCL / LCL computation (measured values)
7. UCL / LCL computation (tolerances)
8. Menu program file management
9. Menu data file management
10.Program START
7.2
Create a learn program (teach-in
method)
1. Press CE key to delete all characteristics that are
stored in the memory
2. Measure the complete workpiece; all functions are
to be carried out automatically after measurement,
e.g. computation functions
3. Press the PROG key and select „Create learn
program“.
4. Enter the tolerances that most frequently occur in
the measuring program as standard tolerances, the
tolerances can be adjusted later see menu point
7.3.
5. Assign the program a name (max. 15 characters).
6. Through program START, the learn program will
start.
If the program name already exits, use the keypad to
enter a new name.
PROG
CE
1. <-_______
2. Learn programming (teach - in)
3. Create new program
4. Change existing program
5. Print out current program
6. UCL / LCL computation (measured values)
7. UCL / LCL computation (tolerances)
8. Menu program file management
9. Menu data file management
10.Program START
Standard tolerances:
_____0.100______
Save program
PRG1
09:57
04/Sep/06
755
TEST1
09:59
04/Sep/06
1175
TEST5
10:18
04/Sep/06
587
New file name
TEST1
PROG
Содержание Digimar 817 CLM
Страница 8: ...8 Mahr GmbH Digimar 817 CLM 1 2 Unpacking 1 2 N I C K L M O P Q...
Страница 10: ...10 Mahr GmbH Digimar 817 CLM 5 350 mm 25 kg 600 mm 30 kg 1000 mm 35 kg...
Страница 11: ...11 Mahr GmbH Digimar 817 CLM 6 45 7 8 90...
Страница 12: ...12 Mahr GmbH Digimar 817 CLM 9 11 G H 10 12 13 M...
Страница 13: ...13 Mahr GmbH Digimar 817 CLM B 3x L M 19 17 18 P D E K N 14 15 16...
Страница 14: ...14 Mahr GmbH Digimar 817 CLM 100 240 V 50 60 Hz I 12 h 21 22 Euro US UK 20 23...