33
Mahr GmbH • Digimar 817 CLM
Description / Sequence
Symbols / Pictures
3.2.3.3
Measuring a bore
Note:
Both the Quick Mode and „measuring a bore“ must be
active. See chapters 3.2.1 and 3.2.2
–
Manually position the probe using the handle eccentrically into the
bore
–
The Quick Mode automatically recognizes the desired measuring
function „measure a bore“ and starts the measuring function
–
The probe moves automatically upwards and makes contact
–
Move the workpiece parallel to the stopping face in order to
determine the reversal point / maximum
–
Acceptance of the reversal point will be confirmed by an acoustic
signal (beep)
–
The probe moves automatically downwards and makes contact
–
Move the workpiece parallel to the stopping face in order to
determine the reversal point / minimum
–
Acceptance of the reversal point will be confirmed by an acoustic
signal (beep)
–
Results (center and diameter) will appear in the display.
Note:
The measurement can be cancelled at any time by pressing any of the
following keys, see chapter 3.4,
or the variable
keys
or by moving the measuring carriage in the opposite direction.