70
Mahr GmbH • Digimar 817 CLM
4.6
Measuring a taper / calculating an angle
On a taper, the angle
α
between its mean line and the base plate is to
be determined. It must be possible to laterally displace the workpiece
by a clearly defined amount. To do this, attach a stop to the base plate
and place a gage block between the stop and the workpiece.
A gage block is required, which is somewhat shorter in length than the
height of the taper.
For the measurement, a probing element with cylindrical styli is
necessary which must be aligned parallel to the base plate.
Procedure
–
Calibrate a cylindrical probe (standard or with a groove),
see chapter 2.1
–
Place the gage block between the workpiece and the stop and
position the probe above the narrow end of the taper.
–
Contact from above (arrow 1)
–
Remove the gage block and slide the workpiece towards the stop
until it makes contact
–
Contact from above (arrow 2)
–
Press the taper function key
The following menu will appear:
–
Using the arrow keys select characteristic A and characteristic B
(use the function keys A and B to accept them)
–
Press taper evaluation and enter the dimension 10.000
–
Confirm – angle will be displayed
Description / Sequence
Symbols / Pictures
ON
OFF
B
A
Содержание Digimar 817 CLM
Страница 8: ...8 Mahr GmbH Digimar 817 CLM 1 2 Unpacking 1 2 N I C K L M O P Q...
Страница 10: ...10 Mahr GmbH Digimar 817 CLM 5 350 mm 25 kg 600 mm 30 kg 1000 mm 35 kg...
Страница 11: ...11 Mahr GmbH Digimar 817 CLM 6 45 7 8 90...
Страница 12: ...12 Mahr GmbH Digimar 817 CLM 9 11 G H 10 12 13 M...
Страница 13: ...13 Mahr GmbH Digimar 817 CLM B 3x L M 19 17 18 P D E K N 14 15 16...
Страница 14: ...14 Mahr GmbH Digimar 817 CLM 100 240 V 50 60 Hz I 12 h 21 22 Euro US UK 20 23...
Страница 152: ......